Purpose
To describe the formulation of 10X Tris-Glycine electrophoretic transfer buffer.
References
Western Blotting of Proteins, Clinical Biotechnology (1989); 1(1):43-49
Materials
- RMS-110 Tris
- RMS-123 Glycine
Equipment
- Stir Plate and stir bar
- Glassware
- pH meter and pH standards
Procedure
- Add the following to 890 ml DI water:
- 30.3 g Tris
- 144.1 g Glycine - Stir to dissolve.
- Standardize pH meter to pH 7 and 10.
- Check pH of solution. It should be between 8.6 and 8.9. DO NOT ADD ACID OR BASE TO ADJUST pH.
- Label with contents, lot number and one year expiration date.